Section 8 Calibration
8-2
8.1 When to Perform Calibration
Calibration is preventive maintenance for preventing degraded performance before it occurs. Even if it appears that the
unit is running normally, you should calibrate the unit to keep in top shape. We recommend you calibrate the unit one to
two times a year. Contact our service department if calibration fails to bring the unit within its rated specifications.
8.2 List of Calibration Equipment
Table 8-1 show the equipment used to calibrate the unit.
Table 8-1 List of Calibration Equipment
Calibration Item
Recommended Equipment
(Anritsu Model Name)
Required Performance *
1
Reference oscillator Host standard
Frequency standard, standard radio wave receiver
Color TV subcarrier
Signal generator
Accuracy
1 × 10
–9
Frequency 1 GHz
Level 0 dBm
*1 :
Some of the performance items that can cover the measurement range are given.