7-3
7.2 List of Performance Test Equipment
7.2 List of Performance Test Equipment
Table 7-1 shows the equipment for performance tests.
Table 7-1 List of Performance Test Equipment
Test Item
7.3.1.1, 7.3.1.2
Continuous frequency
measurement
(Input1, Input2 50 Ω)
7.3.2
Burst wave carrier
frequency measurement
7.3.3
Burst width measurement
7.3.1.3
Continuous frequency
measurement
(Input2 1MΩ)
Recommended Equipment
(Anritsu Model No.)
Power meter (ML4803A)
Power sensor
(MA4701A)
(MA4703A)
(MA4705A)
(MP713A)
Fixed attenuator (20 dB)
*2
(MP721D)
Signal generator
(68247B)
(68259B)
(68269B)
Power meter
Power sensor
Signal generator
Required Performance
*
1
Frequency range
10 MHz to 20 GHz : MF2412B
10 MHz to 26.5 GHz : MF2413B
0 MHz to 40 GHz : MF2414B
Sensitivity –33 dBm to 0 dBm
Frequency range
10 MHz to 20 GHz : MF2412B
10 MHz to 26.5 GHz : MF2413B
10 MHz to 40 GHz : MF2414B
Output level –33 dBm to 0 dBm
Pulse modulator 100 ns
Pulse modulation accuracy ±10 ns or less
Frequency range 10 Hz to 10 MHz
Sensitivity 25mVrms
Frequency range 10 Hz to 10 MHz
Output level 25mVrms
*1 :
Some of the performance items that can cover performance measurement tests have been excluded.
*2 :
Used when testing at –33 dBm.