Section 4 Unit Operation
4-38
4.4 Measuring
4.4.1 Continuous Wave Frequency Measurement via Input1
(Frequency Acquisition Mode = Auto; Level Acquisition Mode = Auto)
Ranges of frequencies that can be measured via Input1 are as follows:
MF2412B: 600 MHz to 20 GHz
MF2413B: 600 MHz to 27 GHz
MF2414B: 600 MHz to 40 GHz
(1) Connecting the input signal
Connect the signal to measure to the Input1 connector on the front panel.
Note:
Do not connect a signal of +10 dBm or higher to the Input1 connector.
(2) Setup
1) Press the [Preset] key to preset this unit. This selects Input1, continuous wave measurement, auto frequency
acquisition, and auto level acquisition.
2) Using the [<] and [>] keys, set a desired frequency measurement resolution.
3) Using the [ ] and [ ] keys, set a desired sample rate.
4.4.2 Continuous Wave Frequency Measurement via Input1
(Frequency Acquisition Mode = Manual; Level Acquisition Mode = Auto)
When the frequency of the input signal is known, you can manually measure the frequency by setting Frequency Acquisi-
tion Mode to Manual and then setting a manual frequency value.
Manual measurement of frequency can start quickly because it is not accompanied by frequency acquisition. It is effective
when measurement cannot be performed due to a spurious signal.
(1) Connecting the input signal
Connect the signal to measure to the Input1 connector on the front panel.
Note:
Do not connect a signal of +10 dBm or higher to the Input1 connector.