EasyManua.ls Logo

Anritsu MF2412B - Page 82

Anritsu MF2412B
177 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4-31
To display the template screen, you must turn the template function On, and then set the various parameters.
Pressing the [Temp] key displays the template setup screen shown in Fig. 4-32. Pressing the [Return to Meas] key in the
displays shown below displays the measurement screen. Fig. 4-31 shows the measurement screen when the template
function is On.
20 000 000 0
00
Hz Template
Upper Limit : 20 000 000 000 Hz
Lower Limit : 19 000 000 000 Hz
Template
[On]
Upper
Limit
Lower
Limit
Indicate
[On]
Fig. 4-32 Template Setup
(1) Menu F1 : Sets the template function to On or Off.
(2) Menu F2 : Sets the upper frequency limit using the front panel numeric keypad. Selecting Upper Limit will highlight
it so that it can be set.
(3) Menu F3 : Sets the lower frequency limit using the front panel numeric keypad. Selecting Lower Limit will highlight
it so that it can be set.
(4) Menu F4 : Turns On/Off the indicator for showing when the measurement frequency strays off of the display screen
(see Fig. 4-4). Select On to display it and Off to not display it.
The upper and lower frequency limit is between 0 Hz and Fmax set in 1 Hz units.
Note, Fmax= 20 GHz.............. MF2412B
27 GHz.............. MF2413B
40 GHz.............. MF2414B
Note :
The Go/No-Go result is stored until the following decision is made.
Measurement state
output
Measurement Measurement Measurement
Pause Pause
Go
No-Go
Go
4.3 Parameters

Table of Contents