EasyManua.ls Logo

Asylum Research cypher - Manifold Connections

Asylum Research cypher
274 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Ch. 14. Gas Handling and Leak Testing Sec. 14.2. Manifold Connections
14.2. Manifold Connections
14.2.1. Scanner Faceplate Removal
1.
Remove the scanner cover:
Grip the cover as shown.
Pull forward. The cover is attached magnetically.
14.2.2. Manifold Cell-side connection
Process for removing the tubing connected on the cell-side of the valve manifold.
1.
Remove the fitting using the tool:
Place the tool on the fitting. The slot in the tool slides over the tubing.
Unscrew until the fitting comes loose. An O-ring should stay attached to the tubing.
If not, you may need to retrieve it from the port with tweezers.
If it feels there is not enough room to complete this step, please follow to the next
step:
DRAFT
Page 175

Table of Contents