12
Aesculap
®
ELAN 4 electro Standard handpiece (1 ring) GA861 to GA864 / ELAN 4 electro
HD handpiece (2 ring) GA865 to GA869
Mechanical alkaline cleaning and thermal disinfection
Machine type: single-chamber cleaning/disinfection device without ultrasound
D–W: Drinking water
FD–W: Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least)
*Recommended: BBraun Helimatic Cleaner alkaline
Connect the connector for motor cable 3 with ELAN 4 electro rinsing
device GB692R.
Place the rinsing device on a wire basket suitable for cleaning.
Connect the rinsing device to the rinsing connection of the rinsing cart.
After the mechanical cleaning/disinfection process:
Remove any residual rinsing water from the applied part using a
compressed-air gun, see instructions for use ELAN 4 rinsing device
and rinsing adapter TA014447 and TA014448.
Check visible surfaces for residues.
Phase Step T
[°C/°F]
t
[min]
Water
quality
Chemicals
I Pre-rinse <25/77 3 D–W -
II Cleaning 55/131 10 FD–W
Concentrate, alkaline:
–pH ~ 13
<5 % anionic surfactant
0.5 % working solution
–pH ~ 11*
III Intermediate rinse >10/50 1 FD–W -
IV Thermal disinfection 90/194 5 FD–W -
VDrying - - - In accordance with the program for the cleaning and disinfecting machine