EasyManua.ls Logo

Barco Encore - Page 198

Barco Encore
404 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
198 Encore Presentation System • User’s Guide
5. Menu Orientation
PIP Adjustment Menu
~ Offset — both cloned images move together with a fixed pixel offset.
For example, if you adjust
H Position, both images move left and right.
If you adjust
V Position, both images move up and down.
Figure 5-52. Clone Offset Mode — Sample Widescreen
~ Mirror — both cloned images move together as if a mirror is positioned
in the middle of the widescreen overlap region. For example, if you
adjust
H Position, both images move towards (and away from) the
center of the widescreen.
Figure 5-53. Clone Mirror Mode — Sample Widescreen
Offset — if “Offset” is selected on the Clone Type line, this field enables you to
set the offset (in pixels). If “
Mirror” is selected, the field is hidden.
In Chapter 7, refer to the “
Using Clone” section on page 308 for operating instructions.
PIP A
Cloned
PIP A
PIP A
Cloned
PIP A
Important
The Clone Setup Menu also appears when the Clone button
is enabled (in the
Layer Functions Section), provided that a
widescreen destination is selected. However, if you press
Clone and the message “Resource Conflict - Cannot Apply
Clone
” appears, the PIP (or Key) is most likely positioned
within the center of the widescreen.
S In a 2 screen blend, the PIP or Key cannot be
positioned such that it sits between screens
1 and 2.
S In a 3 screen blend, the PIP or Key cannot be
anywhere within the screen 2 space.

Table of Contents

Other manuals for Barco Encore