Field Service Manual    December 2003 
    Page 136 of 276 
•  Reconnect the Ssipro Terminal by selecting 'New connection' in the file menu 
when you use a Tera Term window. Select 'serial and COM2'.   Or select 
call(connect) on the shortcut keys when you have a 'Hyperterminal'.  
•  Switch off power and take the 'Ssipro' boards out of the machine. 
•  Move jumper back from position 'J27' to position 'J6'. 
Re-install Ssipro, connect all cables and powerup the SilverWriter.