Field Service Manual December 2003
Page 185 of 276
"LOCKRANGE_ALL" finds the minimum and maximum frequency for which
the PLL locks on DRV4X3. These values have to be stored before
"SKYMAX" is executed.
LTL >2000 4000 LOCKRANGE_ALL
LTL > PROPAR
Enable Quadding in parameter list "option B".
Disable optics calibration in parameter list "option B" and check
whether frequency corrections are zero in parameter list "option M".
Set par 13 (non retuning range) to 10.
At the end of the optic adjustment perform a skymax and save the values.