Field Service Manual December 2003
Page 42 of 276
Led flashes under sw1
With and key you can adjust the value to 2.
This is the Analog output mode (not used)
Press the M key again.
LED flashes under sw2
With and key you can adjust the value to 1.
This is the type of Operation mode.
Press the M key during 5 seconds. The settings are stored.
Adjustment of settings vacuum switch
Press the M and key again
LED under sw1 flashes on and off
With and key you can adjust the level to -300 mbar!!!!!!!!Negative
This is the vacuum level were sw1 will switch on.
Second detection sw2 is not used for suctioncaps.
Press the M key again
LED under +/ flashes on and off
With and key you can adjust the value to 10.
This is the hysteresis.
Press M and key together
With and key you can adjust the last setting to F-0.
Press the M key during 5 seconds. The settings are now stored.
To Zero adjust the sensor
Press the and key together and the sensor will be automatically zeroed.
Sw1-Sw2: ON = no vacuum detected
+/-: - = vacuum + = pressure
Connections
Brown: +V 10.8 30VDC
Blue: common
Black: switch1(pnp 30V 100ma max)
White: switch2 (pnp 30V 100ma max)
Gray: analog (0-5V)
Error messages:
E-1 Overload current is flowing. LED of overloaded SW1 and/or SW2 blinks.
E-2 Pressure is applied during zero adjustment.