Timed Urine and Creatinine Clearance Results
Introduction
UniCel DxC Systems Instructions For Use A13914 System Setup Options
October 2005 Page 4-29
4
Timed Urine and Creatinine Clearance Results
Introduction
When Timed Urine is designated as the sample type for a Sample ID, the Results
report will reflect:
the concentration of the sample aliquot placed on the instrument.
the concentration of the analyte with respect to sample volume and collection time
period.
The DxC 600/800 performs additional calculations to accurately report the
concentration as a function of volume and time.
Calculation Constants and Factors
The sample volume is always expressed in milliliters (mL).
The selectable units for any given chemistry are always expressed as weight per
milliliter (therapeutic drugs only and this would be extremely rare), deciliters or liters.
Therefore, the factors applied to total volume will only be as follows:
For examples of Timed Urine Calculations, refer to the UniCel DxC Synchron Clinical
Systems Reference Manual, CHAPTER 3, Detailed System Setup.
milliliters to milliliters = 1.0
milliliters to deciliters = 0.01
milliliters to liters = 0.001

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Beckman Coulter UniCel DxC Synchron 800 and is the answer not in the manual?

Beckman Coulter UniCel DxC Synchron 800 Specifications

General IconGeneral
BrandBeckman Coulter
ModelUniCel DxC Synchron 800
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals