Bar Code Setup
Introduction
UniCel DxC Systems Instructions For Use A13914 System Setup Options
October 2005 Page 4-33
4
Bar Code Setup
Introduction
The Bar Code Setup option allows the operator to:
Enable/disable sample bar code mode of operation
Enable/disable sample bar code types
Configure sample bar code parameters
Restore defaults
The four sample bar code types that may be used on the DxC are:
Code 39
•Codabar
Interleaved 2 of 5
Code 128
Any combination of bar code types may be enabled.
Setup
From the Setup screen, select <16> Bar Code to enable/disable bar code types, to
further define bar code parameters or to restore defaults
Bar Code Setup may be viewed at any time.
The system must be in Standby or Stopped to make modifications
The Bar Code Setup defaults are:
Bar Code mode: Enabled
All bar code symbologies: Enabled