Loading a Calibrator Diskette
Reagent Load/Calibration UniCel DxC Systems Instructions For Use A13914
Page 5-14 October 2005
Loading a Calibrator Diskette
Serum based calibrators require a calibrator diskette to be loaded on the system. The
calibrator diskette is loaded only one time per lot number.
• To check what lot number is currently loaded on the system, select the Rgts/Cal
icon, then select Options [F6]. Use either the option button or <Pre> and <Next>
to find the calibrator. Then select the calibrator.
• To load a new lot number of calibrator, locate the calibrator diskette in the calibrator
box and follow the procedure below.
Step Action
1 Make sure the DxC is in Standby.
2 Select the Rgts/Cal icon from the menu bar.
3Select Cal Options [F6].
4Select <1> Load calibrator diskette.
5 Verify that the lot number on the diskette is the same as the lot number on
the calibrator bottle. Insert the diskette into the disk drive. Select <OK>.
6 Verify that the lot number on the screen is the same as the lot number on
the calibrator bottle.
7 Remove the diskette from the disk drive.
8
If… Then…
No other diskettes are to be
loaded
Select <Cancel>.
Additional calibrator diskettes
need to be loaded
Select <OK> and repeat Step 5.
The incorrect diskette was placed
in the disk drive
•Select <OK>.
• Obtain the correct diskette.
• Go back to Step 5.