Electronic Maintenance Log
Feature Summary
Maintenance UniCel DxC Systems Instructions For Use A13914
Page 10-4 October 2005
Electronic Maintenance Log
Feature Summary
Maintenance items are grouped together by frequency of performance. Only the
maintenance items appropriate for the particular system and installed hardware are
shown. For example, a system that does not have Closed Tube Sampling (CTS)
hardware will not show the maintenance items for that hardware. On the Scheduled
Maintenance Log screen these groups have been placed under separate tabs labeled:
Weekly
Monthly
Two-Month
Three-Month
Four-Month
Six-Month
Each of these tabs is accessed by using the touchscreen or by using the mouse to
position the cursor over the desired tab and then depressing (clicking) the left button
on the mouse.
In order to provide a pathway directly to the maintenance items that need attention a
yellow DUE indication will also have the following affect on the screen. From any
screen the Utilities icon (wrench) will be yellow. After selecting Utilities the
Maintenance option (option #2) will be yellow. When the Maintenance option is
selected and the Scheduled Maintenance Log screen appears. The tab(s) requiring
action will also be highlighted in yellow.
Selecting the highlighted tabs will reveal the procedures requiring attention
highlighted in yellow. Yellow highlights will also alert the operator when any
maintenance item on the respective tabs is due to be performed. The yellow highlights
appear according to the following schedule.
Weekly - 8 days
Monthly - 31 days
Two-Month - 61 days
Three-Month - 91 days
Four-Month - 121 days
Six-Month - 181 days
Maintenance information will not be lost during normal software version updates but
the maintenance counter will be reset to zero if a complete install of the software is
performed under unusual troubleshooting or system repair situations.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Beckman Coulter UniCel DxC Synchron 800 and is the answer not in the manual?

Beckman Coulter UniCel DxC Synchron 800 Specifications

General IconGeneral
BrandBeckman Coulter
ModelUniCel DxC Synchron 800
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals