Cartridge Chemistry (CC) Reagent Handling System
Introduction
System Description UniCel DxC Systems Instructions For Use A13914
Page 2-18 October 2005
Cartridge Chemistry (CC) Reagent Handling System
Introduction
The Cartridge Chemistry Reagent Handling system is composed of the following
components:
Reagent cartridges
Reagent carousel
Reagent probe assembly
Reagent mixer assembly
Mixer wash cup
The Reagent Handling system is used to transfer reagent from the individual cartridges
to the reaction cuvettes for processing and analysis of the requested chemistry tests.
Reagent Cartridges
Reagent cartridges are single use, recyclable plastic containers that house the
individual liquid reagent components necessary to perform a chemistry test. The
reagent carousel is capable of storing 59 cartridges on board.
Figure 2.11 CC Reagent Cartridge
1. A compartment
2. B compartment
3. C compartment