Summary of Hazards
Hazardous/Biohazardous Substances Hazards
UniCel DxC Systems Instructions For Use A13914 General Information
October 2005 Page 1-5
1
Hazardous/Biohazardous Substances Hazards
When handling a spill of blood or other potentially hazardous substances, clean up the
spill by using a 10% bleach solution, or use your laboratory decontamination solution.
Then follow your laboratory procedure for disposal of hazardous materials. If the
UniCel DxC system needs to be decontaminated, call your Beckman Coulter Service
Representative for assistance.
ISE Module Hazards
When lowering the ISE module, keep fingers clear to avoid a pinching hazard.
Moving Parts Hazards
Do not place hands near any moving parts while the system is in operation.
No Foam Reagent Pressurization Hazard
The No Foam container is pressurized during system operation and must be properly
depressurized prior to servicing to avoid sudden depressurization and potential
exposure of the skin or eyes to the No Foam solution. To release the air pressure,
disconnect the white quick connector located at the supply side of the container. If
inhaled, move exposed individual to fresh air. If skin or eye contact with the solution
occurs, flush the affected area thoroughly with water for at least 15 minutes. In both
cases seek medical attention. Refer to the Antifoam Material Safety Data Sheets
(MSDS) for additional information.
Power Cord Hazards
Only use a three-pronged power cord to connect the instrument to a matching three-
wire grounded outlet. Do not use an adapter to connect the power plug to a two-
pronged outlet.
Service Procedures Hazards
Disconnect the power cord when performing service procedures such as replacing
electronic or mechanical components.
The reagents and other chemical preparations used with the system will not normally
cause adverse reactions; however, those persons with sensitive skin should wear
protective rubber gloves before attempting to work with those substances.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Beckman Coulter UniCel DxC Synchron 800 and is the answer not in the manual?

Beckman Coulter UniCel DxC Synchron 800 Specifications

General IconGeneral
BrandBeckman Coulter
ModelUniCel DxC Synchron 800
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals