Calibrator Set Point Modifications
Modifying Calibrator Set Points
UniCel DxC Systems Instructions For Use A13914 Reagent Load/Calibration
October 2005 Page 5-31
5
5 Move the cursor to the set point to be modified and type in the desired
value.
If a previously modified set point is to be returned to the original default
value for that calibrator or chemistry, select desired chemistry, move
cursor to the applicable set point and select <Restore Defaults>.
The default values have more significant figures than shown on the screen.
Therefore, select <Restore Defaults>, rather than typing the default
values.
6Select <Close> to exit the screen.
Calibration status changes to Cal Required. (Status changes only for those
containers with the same lot number.)
A flag indicating that specific set points have been modified will appear in
the Instrument Code section on the results report.
Step Action, continued
(2 of 2)