Processing Parameters
Usable Result Range
UniCel DxC Systems Instructions For Use A13914 User Defined Reagents
October 2005 Page 9-15
9
Usable Result Range
Usable result range specifies the analytical range of the reagent. Results will be
suppressed and flagged as out-of-instrument range low (OIR-LO) or out-of-instrument
range high (OIR-HI), respectively, if these ranges are exceeded.
Table 9.25 Reaction Read Times
Allowable Entries
Single Cycle Test
Double Cycle Test
a
a
Overall system throughput is decreased. Cuvettes next to the test cuvette are left empty to prevent the test cuvette from being
washed in the first cycle.
Start Read 1 to 1704 seconds 1 to 1604 seconds
End Read 1 to 1720 seconds 721 to 1620 seconds
Table 9.26 Usable Result Range
Allowable Entry
Lower Limit 0.000 to 99999.999
Upper Limit 0.000 to 99999.999