Wavelength Selection
Summary
UniCel DxC Systems Instructions For Use A13914 User Defined Reagents
October 2005 Page 9-19
9
Summary
A secondary wavelength closer to the primary wavelength reduces noise and
increases accuracy.
Obtaining the maximum absorbance difference increases sensitivity.
Avoid secondary wavelengths with peak spectral interferences within the primary
spectral curve.
Select a secondary wavelength that produces a net absorbance that is close to the net
absorbance of the primary wavelength.