Monthly Maintenance
Replace CTS Blade (Sarstedt S-Monovette Type CTS Option only)
UniCel DxC Systems Instructions For Use A13914 Maintenance
October 2005 Page 10-33
10
Material Required:
Medium straight-edge screwdriver
Replacement Procedure
In the following procedure, the old blade is removed in Steps 1–5 and the new blade is
installed in Steps 6–11.
Step Action
1 To access this procedure, select:
the Utils icon from the menu bar,
<2> Maintenance,
<1> 1-Blade Thick CTS Blade Replacement.
To start the procedure, select <Continue>.
OR
To exit this screen, select <Cancel>.
2 The instrument prepares for blade replacement. Then the Maintenance
Confirmation Window appears.
(1 of 4)