Operational Conditions
Water Requirements
System Description UniCel DxC Systems Instructions For Use A13914
Page 2-4 October 2005
Water Requirements
Other System Specifications for IEC-1010 Compliance
Table 2.5 Water Requirements
Item Specification
Flow Rate 0.6 L/min peak flow rate
16 L/hr, minimum continuous flow rate
Temperature +15°C to +25°C
Water quality NCCLS Type II deionized water, except for total
bacteria count < 10 cfu/mL
Water pressure Deionized water entering the system must be 30–90
psig.
Table 2.6 IEC-1010 Specifications
Item Specification
Pollution Degree 2
EN55011 Meets Class A
Maximum Sound Pressure ≤ 65 dBA average over 8 hours with covers down at 1
meter away from the instrument at 25°C
Maximum Leakage Current DxC 600: 222 µA at 240V, 50Hz
DxC 800: 240 µA at 240V, 50Hz