EasyManuals Logo

Beckman Coulter UniCel DxC Synchron 800 User Manual

Beckman Coulter UniCel DxC Synchron 800
510 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #310 background imageLoading...
Page #310 background image
Six-Month Maintenance
Replace Glucose Sensor
Maintenance UniCel DxC Systems Instructions For Use A13914
Page 10-64 October 2005
Sensor Replacement and Stir Bar Cleaning
Follow the steps below to replace the sensor. After replacement, perform a Sensor
Calibration, then calibrate Glucose.
Step Action
1 Drain the reagent from the MC cup by selecting:
the Utils icon from the menu bar,
<2> Maintenance,
<4> Cup Maintenance.
2Select Glucose cup to be drained in preparation for maintenance.
Select <Drain>. This will drain the cup of reagent so that maintenance
may be performed,
OR
Select <Cancel> to immediately return to the Maintenance Main Menu
without draining the cup.
3 Proceed to Sensor Replacement and Stir Bar Cleaning.
NOTICE
When a new sensor is installed, if needed, the reaction cup and/or reagent lines
can be cleaned at this time. Follow the Four-Month Maintenance procedures
found in the Four-Month section of this chapter.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Beckman Coulter UniCel DxC Synchron 800 and is the answer not in the manual?

Beckman Coulter UniCel DxC Synchron 800 Specifications

General IconGeneral
BrandBeckman Coulter
ModelUniCel DxC Synchron 800
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals