As-Needed/As-Required Maintenance
Cleaning the ISE Drain
UniCel DxC Systems Instructions For Use A13914 Maintenance
October 2005 Page 10-95
10
4 Loosen the screw at drain top (1). Gently grasp the ISE drain top and
remove the top. Do not lose the attached metal drip screen (2). (The top
may require slight wiggling back and forth to remove.)
Figure 10.69
5 Carefully set aside the drain top (with tubing #22 attached).
6 Remove the drain tube (1) from the drain.
Figure 10.70
7 Using four to five cotton swabs soaked with deionized water, clean the
inside of the lower drain and the top of drain.
Step Action, continued
(2 of 3)