Sample Handling System
1-Blade Thick CTS (Closed Tube Sampling) Cap Piercer Assembly (optional)
UniCel DxC Systems Instructions For Use A13914 System Description
October 2005 Page 2-11
2
1-Blade Thick CTS (Closed Tube Sampling) Cap Piercer Assembly (optional)
This optional 1-Blade Thick CTS Cap Piercer assembly pierces thick-stoppered tubes
allowing the sample probe access to the sample without the need for cap removal.
Tubes with caps must be loaded in racks that have had the size correctly defined and
that are not Reserved (Reserved Racks = No CTS). Both open and closed tubes can be
in the same sample tube rack.
Refer to "Symbols and Labels" in CHAPTER 1, General Information, for a description
of the CAUTION labels for the Cap Piercer.
CAUTION
This Cap Piercer contains a razor sharp blade assembly.
CAUTION
Do NOT use this Cap Piercer assembly with foil-capped tubes. This would
give level sense errors.
Table 2.8 Tubes Validated for Closed Tube Sampling
Cap Piercer
Configuration
Tube Type Tube Size
1-Blade Thick CTS
Becton Dickinson VACUTAINER
®a
with HEMOGARD™
a
a
All trademarks are the property of their respective owners.
13 × 75 mm
13 × 100 mm
16 × 100 mm
Greiner VACUETTE
®a
13 × 75 mm
13 × 100 mm
Sarstedt S-Monovette
®ab
b
This tube requires a special 5.5 mL rack (P/N A18642).
75 × 15 mm
92 × 15 mm

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Beckman Coulter UniCel DxC Synchron 800 and is the answer not in the manual?

Beckman Coulter UniCel DxC Synchron 800 Specifications

General IconGeneral
BrandBeckman Coulter
ModelUniCel DxC Synchron 800
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals