LB 4700
8 Maintenance and Repair
56926BA2, Rev. 02, 04/2019
75
8.7 Replacing the Scintillator/Multiplier Combination
j
DANGER
Danger to life from electric shock!
u The installation may only be carried out by a qualified electrician.
u Please adhere to the relevant safety regulations.
u Installation/maintenance may only be carried out if the device has been de
-
energised.
u Only open the device when free of voltage.
In case of an electric shock, carry out first aid measures and immediately call an
emergency service.
WARNING
Danger to life by explosion!
u
If the detector is located in a potentially explosive atmosphere, it is essential
to observe the
corresponding instructions and safety instructions in the
safety manual.
1. Remove the electronics module through the action steps 1 - 7 as described in
the chapter 8.5.1.
2. Insert the new crystal/multiplier combination in the socket. Please observe the
coding lug.
3. Insert the electronics module again, see chapter 8.5.1.
4. Check the measuring function. Should there be discrepancies, perform a new
calibration (see operating instructions of the associated evaluator device).