LB 983 NightSHADE Operating Manual Trouble Shooting
107
7.1 Hardware
7.1.1 Door status always open
If the door status is always open, the reason could be a disaligned door sensor. The door
senssor is located at the upper right front side of the dark box. To adjust the door sensor refer to
8.2.3
7.1.2 Fluorescence lamp not working
The built in halogen fluorescence excitation light source has a lifetime of approx. 30h. When the
lamp is no longer working replace the lamp. Refer to 8.2.4