EasyManuals Logo

Biotek BioStack 3 User Manual

Biotek BioStack 3
127 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #90 background imageLoading...
Page #90 background image
68 | Chapter 4: Maintenance
5. Apply a small amount of new lubricant to all surfaces of the linear ways,
spreading thin layers and working the grease into the surfaces with a
toothbrush.
l Ensure that you lubricate the sides of the linear ways, since ball bearings
travel on these surfaces.
6. Return the BioStack to its upright position and reconnect its power supply.
7. Turn on the BioStack and run a Verify Test to confirm that moving the BioStack
did not compromise its alignment with the interfacing instrument. (Refer to the
instrument-specific PDF file.)
l If the Verify Test passes, the BioStack is ready to resume normal operation.
l If the Verify Test fails, realign the gripper with the plate carrier using the
Alignment Utility and rerun the Verify Test. If the test continues to fail,
contact BioTeks Technical Assistance Center.
BioStack™ Microplate Stacker

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Biotek BioStack 3 and is the answer not in the manual?

Biotek BioStack 3 Specifications

General IconGeneral
BrandBiotek
ModelBioStack 3
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals