95
Follow the steps below to perform the current calibration:
1. Connect the DMM to the output of the power supply as shown
in the figure below:
Figure 5.2 – Current Calibration Diagram
2. Set the DMM to a DC current measurement. Select the
CURRENT parameter in the CAL menu and press . The
following will be displayed:
The P: 1/4 signifies there are 4 points to be calibrated, and the
current calibration point is 1.
3. Input the current reading from the DMM in the METER
parameter by using the numeric keypad and press .