Default Icon
192 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3. Refer to Figure 1-1. Check that the oil-level is between the MAX and MIN marks on the bezel
of the HCMB oil-level sight-glass (5):
•
If the oil-level is above the MAX mark, drain excess oil from the pump as described in
Section 6.10 until the oil-level is correct.
•
If the oil-level is below the MIN mark, fill the pump with oil until the oil-level is correct, as
described in Section 6.10.
4. Refit the right-hand side enclosure panel: refer to Section 3.17.
6.8 Check the HCDP pump oil-level
WARNING
Do not remove the oil filler-plug while the pump is operating. If you do, hot
oil may be ejected from the pump.
WARNING
The majority of synthetic oils/grease can cause inflammation of the skin
(dermatitis). Safety precautions must be taken to prevent prolonged skin
contact with these substances. Use of suitable protective gloves and clothing
along with respirator is recommended if contact with substances is
anticipated.
CAUTION
Do not check the oil-level when the iH system is operating. The correct oil-level may not be
shown when the pump is operating.
CAUTION
Ensure that the oil-level in the HCDP pump is correct. If the pump oil-level is incorrect,
pump performance may be affected and the pump may be damaged.
Use the following procedure to check the oil-level.
1. Shut down the iH system (refer to Section 5.10) and allow it to cool.
2. Remove the right-hand side enclosure panel from the iH system: refer to Section 3.4.