2.5 Lubrication
Note: BOC Edwards Material Safety Data Sheets for the oils referenced in the following sections are
available on request. (Refer to Section 1.1 for contacts).
2.5.1 HCDP pump
Gearbox
Oil charge 0.75 litres (typical)
Grade of oil SAE 40
ISO viscosity grade 150
Recommended perfluoropolyether oils Fomblin 25/6, Krytox 1525, Drynert 25/6
2.5.2 HCMB pump (iH600/iH1000 only)
Gearbox
Oil charge 0.73 litres (iH600), 0.78 litres (iH1000)
Grade of oil SAE 40
ISO viscosity grade 150
Recommended perfluoropolyether oils Fomblin 25/6, Krytox 1525, Drynert 25/6
2.6 Electrical data
Electrical supply
Supply voltage 200-208/380-415 V at 50 Hz, 3-phase
200-208/230/460 V at 60 Hz, 3-phase
Voltage tolerance ±10% (except for 208 V and 415 V at
50 Hz, which are +6% and -10%)
Installation Category Category 2 (IEC 664)
Full load current ratings See Table 2-2
Fuse types and ratings
F1
✝
30 A, 600 V, Class J
F2
✝
30 A, 600 V, Class J
F3
✝
30 A, 600 V, Class J
F4
✝
20 A, 600 V, Class J
F5
✝
20 A, 600 V, Class J
F6
✝
20 A, 600 V, Class J
F7 1.6 A, 250 V
F8 1.6 A, 250 V
F9 3.15 A, 250 V
Emergency stop switch rating 24 V a.c./d.c., 5 A max
External emergency stop switch
Rating 24 V a.c., 5 A
Load 700 mA (inductive)
✝ These fuse rating are included for information only. You cannot change these fuses;
a BOC Edwards service engineer must change these fuses
2-14 iH Dry Pumping Systems