Alarm message Advisory message Meaning/recommended action
ALARM 17601
SEE MANUAL The inverter on an iH1000 system has
experienced a significant power supply
problem and the HCMB pump has been
stopped, or the HCMB pump in the iH1000
system has stopped because of a mechanical or
electrical fault. You must manually restart the
HCMB pump (refer to section 5.13). If the
alarm condition persists after several attempts
to restart the HCMB pump, contact your
supplier or BOC Edwards for advice.
Table 6-2 - Alarm messages (continued)
Message Meaning
CONTROL failed Press
CANCEL
You have requested that the Pump Display Terminal takes
control of the iH system, but the attempt to take control has
failed. Press the CANCEL button to return to normal display.
CONTROL REMOTE
Press CANCEL
(a) You have requested that the Pump Display Terminal takes
control of the iH system, but another module still has control.
Press the CANCEL button to return to normal display, release
control from the module which currently has control and then
try again.
(b) You have requested some action, but the Pump Display
Terminal does not have control of the iH system. Press the
CANCEL button, take control and then try again.
PUMP failed to
respond (CANCEL)
(a) You have requested that the iH system is switched on or off,
but the pump(s) has failed to start or stop: press the CANCEL
button to cancel the request, identify and rectify the cause for
the failure of the pump(s) to start or stop, then try again.
(b) You have tried to open or close one of the gate valves (if
fitted), but the valve has failed to respond. Identify and rectify
the cause for the failure of the valve to open or close, then try
again.
(c) A valve (for example, a setpoint) has been requested from
the iH system, but the iH system has not been responded.
(d) The emergency stop switch has been pressed and released:
restart the pumping system.
PUMP RUNNING
Press CANCEL
You have requested that the iH system is switched on, but it is
already on. Press the CANCEL button to return to normal
display.
6-28 iH Dry Pumping Systems
Table 6-3 - Fault messages