check the connections and phase measure
measurement readings you get are indeed
ments if this is not the case to make sure the
correct.
If it is necessary to adjust the pots, see Table 6-6 for the corresponding pot designators. The
top e
the
boa hase B/C board (7000-722).
erence
cover has to b removed to access these pots. They are located along the top edge of
controller board(s). For Series I MX units, adjustments for phase A are on the phase A
rd (7000-721), adjustments for phase B and C are on the p
For Series II MX units, all adjustments are on the 7003-718 controller board. The ref
designators are the same on either series controller.
12. Repeat steps 2 through 11 for phase B and C. (except on single phase only models).
MX Series I R resistive Xl inductive
Phase / Board #
Phase A (7000-721) R121 R122
Phase B (7000-722) R112 R111
Phase C (7000-722) R114 R115
MX Series II R resistive Xl inductive
Phase / Board #
Phase A (7003-718-2 / -4) R121 R122
Phase B (7003-718-2) R112 R111
Phase C (7003-718-2) R114 R115
Table 6-6: Programmable Z adjustment pots
Definitions:
V
NL
= Measured RMS voltage under no load.
V
L
= Measured RMS voltage under load
I = Measured RMS current.
F = Source frequency (50 Hz).
= Ph angle shift between load and no load conditions. Record phaase se angle
from phase meter under NL and L condition and determine phase shift.
Formulas to calculate R and L component of output impedance:
R = ( V
NL
* cos() - V
L
) / I
X
L
= ( V
NL
* sin()) / I
L = X
L
/ (2 * Pi * F)
Table 6-7: Formulas to calculate R and L
MX Series 175