User Manual - Rev AJ California Instruments
9.5.4.3 RUNNING THE TEST
The test seq
appropriate
uence used is a function of the EUT class. The end user needs to determine the
class of EUT. The test protocol is document in the IEC 61000-4-13 test standard.
For reference, the test flow charts are shown here. It is assumed that the end user has a copy of
the actual test standard however.
Start - Class 1/2
Determine appropriate test
igurationconf
Perform
"Harmonic
Flatcurve and Overswing
8.2.1 Test
combination"
Any functional
anomalies?
No No
Class 2
required?
curve' req
'Meister
uired?
Any functional
anomalies?
Any functional
anomalies?
Any functional
anomalies?
Test Completed
R
Test Completed
PASS
Record Results
FAIL
ecord Results
No
Yes
Yes
No
Perform 8.2.2 Test
"Frequency Sweep"
Perform 8.2.4 Test
"Meister Curve"
Perform 8.2.3 Test
"Individual Harmonics/
Interharmonics"
Yes
Yes
No
Yes
Yes
No
8: IEC 61000-4-13 Test Flowchart Class 1 and 2 Figure 9-2
MX Series
220