User Manual - Rev AJ California Instruments
To run the test, the IEC 1000-4-13 mode must be selected. Refer to paragraph 4 for access to
the screen. The following conditions have to be met before running the test:
1. Prior to the test, set the nominal voltage, frequency, and phase angle. The frequency must
be 50 or 60 Hz. Function m
2. Select EUT CLASS 2, 3 or USER. The default parameters for the USER class are identical
. If USER class is selected, the level and phase parameters can be
4. If it is desired to run all tests, select ALL from the RUN parameter with the shuttle knob and
FCURVE, OSWING, SWEEP, HARM, IHARM
field by rotating the shuttle.
6. If it is desired to run a specific group, the group must be selected from the group screen.
Also, the RUN GROUP must be selected prior to the start of test.
7. To run an individual range, applicable for sweep and individual harmonics only, the desired
range and the RUN RANGE mode must be selected prior to the start of the test. In case the
harmonics group is selected, RUN RANGE will run the harmonics from the second
gure 9-25.
8. To run individual harmonics or individual inter harmonics, set the harmonic number or the
inter harmonics frequency and select RUN POINT prior to running the test.
ust be set to sine wave.
to those for class 3
edited.
3. Set the STATE to ON. It will take a few seconds to initialize data required to perform the
test.
press the ENTER key. The tests will run in the following sequence:
Note: The test for class 3 will take about 25 minutes, class 2 slightly less.
5. While the test is running, the ABORT and PAUSE selection is accessible from the RUN
harmonics to the harmonics number specified by the HARM NO field. See Fi
MX Series
222