User Manual - Rev AJ California Instruments
Number Message String Cause Remedy
executed
programming frequency when in
with mode of operation such as
DC mode.
-201 "Invalid while in local" Command
ot in rem state before issuing GPIB commands.
issued but unit Put instrument in remote state
is n ote
-203 "Command protected" Command Some commands are supported is locked out
by the unit but are locked out for
protection of settings and are
not user accessible.
-210 "Trigger error" Problem wi
system.
th trigger Unit could not generate trigger
for transient execution or
measurement.
-211 "Trigger ignored" Trigger request has been Trigger setup incorrect or unit
ignored. was not armed when trigger was
received. Check transient
system or measurement trigger
system settings.
-213 "Init ignored" Initialization
en ignore
request has Unit was told to go to armed
be d state but was unable to do so.
Could be caused by incorrect
transient system or
measurement acquisition setup.
-220 "Parameter error" Parameter not allowed. Incorrect parameter or
parameter value. Check
programming manual for
allowable parameters
-221 "Setting conflict" Requested
to program a DC offset while in
AC mode
setting conflicts Check other settings. E.g. trying
with other setting in effect.
-222 "Data out of range" Parameter outside of
wable r
Check programming manual for data
allo ange. allowable parameter values
-223 "Too much data" More data received than
expected
Check programming manual for
number of parameters or data
block size
-224 "Illegal parameter
value"
Pa
supported
rameter value is not Check programming manual for
correct parameters
-226 "Lists not same length" One or mor
programme
length.
e transient lists
All lists must be of same length
d has different
or transient cannot be compiled
and executed.
-241 "Hardware missing" N/A N/A
-254 "Media full" No storage
save settings or data. make room.
space left to Delete other settings or data to
-255 “Directory full” To
dir
o many w
ectory en from waveform memory to make
aveform Delete one or more waveforms
tries
room.
-256 “File name not found” Wa
dir
veform
ectory
requested not in Check waveform directory for
waveform names present.
-257 “File name error” Incorrect fil
definition.
ename Too many or non ASCII
characters used in waveform file
-283 “Illegal variable name” Variable name illegal. Use ASCII characters only
-300 "Device specific error" Hardware related error Check hardware for proper
MX Series
240