67
11. Wait for the balance to weigh the sample.
Start Weight
SAMPLE ID: XXXX
SAMPLE WT: XXX
Press NEXT key to weigh next sample,
or press SELECT for run screen
12. Press the SELECT key.
XXXXXXXX
RAMP XX C
XX:XX:XX Set Point
DWELL XXX C
XX:XX:XX
Status: Paused
WARNING
To prevent the possibility of serious burns, ensure that insulated
gloves and protective gear as outlined in the users safety program
are worn and metal tongs are used to insert and remove sample con-
tainers from the ashing furnace.
13. Open the instrument door. Remove the furnace door and place it in the furnace
door holder.
14. Using metal tongs, place the crucible(s) containing the sample(s) into the ashing
furnace.
Note: When the furnace door is removed, a beeping alarm will sound until the
furnace door is installed.
Note: When the furnace door is positioned in the furnace door holder, the instrument
door cannot be closed.
15. Install the furnace door and close the instrument door.
16. Press the START key to activate microwaves.
XXXXXXXX
RAMP XX C
XX:XX:XX Set Point
DWELL XXX C
XX:XX:XX
Status: Stage 1