68
17. The programmed ramp time for Stage 1 begins counting down. When the ramp
time is complete, the dwell time for Stage 1 will begin counting down. Ramp
time and dwell time will continue through each programmed stage. When the
nal dwell time is complete, a series of audible beeps will be heard, signaling
the end of the method.
Note: If insuf cient ramp time is programmed, and the set-point temperature is not
achieved, an audible beep will be heard, and the display will alert the operator that
conditions have not been met.
Test Parameters Not Met!
BACK - ABORT NEXT - CONTINUE
Note: Press the NEXT key to continue the method although conditions were not met.
Press the BACK key to abort the method so that parameters can be edited.
WARNING
To prevent the possibility of serious burns, ensure that insulated
gloves and protective gear as outlined in the users safety program
are worn and metal tongs are used to insert and remove sample con-
tainers from the ashing furnace.
18. Open the instrument door. Remove the furnace door and place it in the furnace
door holder.
19. Using metal tongs, remove the crucible(s) containing the sample(s) from the
ashing furnace.
20. Install the furnace door and close the instrument door. Press the START key.
21. Press the “Balance” key.
Edit Sample ID
ENTER NEW ID
FINAL WEIGHT
ERASE ID
SELECT ANALYSIS ASH
22. Using the arrow keys, highlight Final Weight. Press the SELECT key.