EasyManuals Logo

Chroma 63600-2 User Manual

Default Icon
258 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #241 background imageLoading...
Page #241 background image
Verification
7-7
7.3.2.3 Checking Low ohm Range
A. Press till VFD shows CR and press to light up the L range LED
indicator.
B. The current shunt range is 250A. Press to input the resistance listed in Table
7-10 . Press to enable load, and see value of DMM(V) to adjust the value of DC
source same as the setting value for testing model before, and waited for 30 seconds,
record the voltage that passes through the load input terminals DMM (V) and the shunt
current reading DMM (I). Calculate the values of the resistance via DMM (V)/DMM (I).
Check the values to fit the specification.
Table 7-10
Model Name
Resistance
Setting
Input Value of
DC Source
Appropriate Conductance (S)
Max.
Min.
63630-600-15
270Ω
80V
0.02371S
0S
0.133Ω
2V
7.546316S
7.491278S
63640-150-60
60Ω
12V
0.067S
0S
0.2Ω
8V
5.072S
4.928S
63630-80-60
30Ω
6V
0.233367S
0S
0.015Ω
0.8V
66.9333S
66.4S
63610-80-20
80Ω
6V
0.087513S
0S
0.04Ω
0.8V
25.1S
24.9S
63640-80-80
20Ω
6V
0.32505S
0 S
0.01Ω
0.8V
100.375S
99.625S
7.3.3 CV Mode Verification
This test verifies if the voltage programming and reading value on the front panel display are
within specifications when the module is operating in CV mode. For each DMM (V) reading,
the front panel display of voltage should be equivalent to:
Load module reading in volts = DMM (V) reading in volts ± inaccuracy.
7.3.3.1 Checking High Voltage Range
A. Connect the Load module, DC source, DMM and current shunt as Figure 7-1 shows.
Use DMM (V) to measure the voltage passing through the module’s input terminal. Be
careful in making connections so that contact resistance voltage drop will not affect the
readings.
B. Press till VFD shows CV and press to light up the H range LED
indicator.
C. Press to set load voltage and press to set limit current. The DC Source
voltage output and limit current settings are based on the voltage/current values listed in
Table 7-11.
D. Next, press to enable the load and wait for 30 seconds to record the voltage
passing through the negative input terminal.
MODE
RANGE
EDIT
LOAD
MODE
RANGE
EDIT
DATA
LOAD

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Chroma 63600-2 and is the answer not in the manual?

Chroma 63600-2 Specifications

General IconGeneral
BrandChroma
Model63600-2
CategoryPower Supply
LanguageEnglish

Related product manuals