[ADVance:]USER:WAVeform:CLEar?
Type: Channel-Specific
Description: Clear the waveform specified.
Setting Syntax: [ADVance:]USER:WAVeform:CLEar?<space><NR1>
Setting Parameters:<NR1>, 1 ~ 10, Resolution = 1, Unit = None, 1~10:Waveform 1~10
Setting Example: ADV:USER:WAV:CLE? 3
Query Syntax: None
Return Parameters:<NR1>, 0:ok 1:error
Query Example: None
5.3.2.3 CHANNEL Subsystem
:CHANnel
:
ID
:LOAD
SP
<NRf
+
>
?
SP
MAX
MIN
?
<NR
1>
<
AARD>
<NR
1>
<
NR1
>
:ACTive
SP
OFF
/ 0
ON / 1
?
<
CRD>
CHANnel[:LOAD]
Type: Channel Specific
Description: Selects a channel of which the coming channel-specific command
will be received and executed.
Setting Syntax: CHANnel[:LOAD]<space><NRf+>
Setting Parameters:63600-1:1 ~ 2 63600-2:1 ~ 4 63600-5:1 ~ 10
Setting Example: CHAN 1 Set the channel to "1".
CHAN MAX Set the channel to "10".
CHAN MIN Set the channel to "1".
Query Syntax: CHANnel[:LOAD]?[<space><MAX | MIN>]
Return Parameters:<NR1>, 63600-1:0 ~ 2 63600-2:0 ~ 4 63600-5:0 ~ 10
[Unit = None]
Query Example: CHAN?
CHAN? MAX
CHAN? MIN
CHANnel:ACTive
Type: Channel Specific
Description: Enables or disables the load module.
Setting Syntax: CHANnel:ACTive<space><CRD | NR1>
Setting Parameters:<CRD | NR1>, OFF(0), ON(1)
Setting Example: CHAN:ACT 1 Enables the load module.
CHAN:ACT OFF Disables the load module.
Query Syntax: CHANnel:ACTive?
Return Parameters:<CRD>, OFF, ON [Unit = None]