Local Operation
4.7.8 Setup of Digitizing Function
To record the transient voltage and current waveforms, the 63600 series offer a digitizing
function for recording the transient waveforms. It is very convenient to record the information
via this function.
In the page of system configuration, turn the Rotary knob to change the display value to 9,
then press key into Digitizing Function edit page of system configuration.
Trig
_point
Sampling point
Points
Voltage or
Current
Sampling Time
Sampling point≤
4 09 6
point
Trig signal
Description of Parameter Setting:
Set the Sampling Time. Set the interval of sampling time. The range is from 2μs to 40ms,
and the resolution is 2μs. The default setting of Sampling Time is 40ms.
Set the Sampling Point. Set the sampling points. The range is from 1 to 4,096 points. The
default setting of Sampling Point is 4,096 points.
Set the Trigger point. Set the Trigger point of Digitizing Function. The range is from 1 to
4,096 points. The default setting of Sampling Point is 2,000 points.
Set the Trigger Source of Digitizing Function. Load ON, Load OFF, TTL (External trigger,
TRIG_DIGI signal), BUS trigger, and Manual trigger could be chosen to be the Trigger
Source. The default setting of Trigger Source is Load ON.
LOAD ON: It triggers at Load on
LOAD OFF: It triggers at Load off.
TTL: It triggers external TTL (Pin No. 7 of the System I/O Port → TRIG_DIGI)
BUS: It executes the command DIGitizing:TRIGger ON for trigger.
MANUAL: Press EDIT+ENTER on the Module front panel at the same time to
trigger it.
Procedures:
STEP 1: Setting the parameters