EasyManua.ls Logo

Chroma 63600-2 - MODE Subsystem

Default Icon
258 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Operation
5-69
5.3.2.11 MODE Subsystem
:MODE
SP
CCM CCH
CRL
?
<CRD
>
CRM
CRH
CVL
CVM CVH
CPL
CPM
CPH
CZL
CZM CZH
CCDL
CCDM CCDH
CCFSL
CCFSM CCFSH
TIML TIMM
TIMH
SWDL
SWDM SWDH
OCPL OCPM OCPH
PROG
CCL
MPPTL MPPTM MPPTH
UDWL UDWM UDWH
MODE
Type: Channel-Specific
Description: This command sets the operational mode for the electronic load.
Setting Syntax: MODE<space><NRf>
Setting Parameters:<CRD>, CCL, CCM, CCH, CRL, CRM, CRH, CVL, CVM, CVH, CPL,
CPM, CPH, CZL, CZM, CZH, CCDL, CCDM, CCDH, CCFSL,
CCFSM, CCFSH, TIML, TIMM, TIMH, SWDL, SWDM, SWDH,
OCPL, OCPM, OCPH, PROG, MPPTL, MPPTM, MPPTH, UDWL,
UDWM, UDWH
Example: MODE CCL Set CC mode of low range.
MODE CCH Set CC mode of high range.
MODE CCDL Set CC dynamic mode of low range.
MODE CCDH Set CC dynamic mode of high range.
MODE CRL Set CR mode of low range.
MODE CRH Set CR mode of high range.
Query Syntax: MODE?
Return Parameters:<CRD>, CCL, CCM, CCH, CRL, CRM, CRH, CVL, CVM, CVH, CPL,
CPM, CPH, CZL, CZM, CZH, CCDL, CCDM, CCDH, CCFSL,
CCFSM, CCFSH, TIML, TIMM, TIMH, SWDL, SWDM, SWDH,
OCPL, OCPM, OCPH, PROG, MPPTL, MPPTM, MPPTH, UDWL,
UDWM, UDWH
Query Example: MODE?

Table of Contents