8. Start-up & programming
Start-up & programming
Error Error text #Comment Outputs Input
No. Remedy text status status
1 I1 -
Power on
OK Power on has happened Active Active
2 I2 -
Add-on module
Applied A new module has been applied to the system Active Active
3 I3 -
Add-on module
Install An add-on module is defect or has been removed.
This can be an internal add-on module Active Active
4 I4 -
Param. corrected
OK A less vital parameter in the converter has been re-
placed by its default value Active Active
20 W20 -
Totalizer 1
Reset manually During initialisation the check of the saved totalizer
value has failed. It is not possible to rely on the
saved totalizer value anymore. The totalizer value
must be reset manually in order to rely on future
readings Active Active
20 W20 -
Totalizer 2
Reset manually During initialisation the check of the saved totalizer
value has failed. It is not possible to rely on the
saved totalizer value anymore. The totalizer value
must be reset manually in order to rely on future
readings Active Active
21 W21 -
Pulse overflow
Adj. pulse settings Actual flow is too big compared with pulse width and Reduced
volume/pulse pulse width Active
22 W22 -
Batch timeout
Check installation Duration of batching has exceeded a predefined Batch out-
max. time put on zero Active
23 W23 -
Batch overrun
Check installation Batch volume has exceeded a predefined maximum Batch out-
overrun volume put on zero Active
24
W24 - Batch neg. flow
Check flow direction Negative flow direction during batch Active Active
30 W30 -
Overflow
Adj. Q
max.
Flow is above Q
max.
settings Max. 120 % Active
31 W31 -
Empty pipe
Pipe is empty Zero Active
40 P40 -
SENSORPROM
â
Insert/change SENSORPROM
â
unit not installed Active Active
41 P41 -
Parameter range
Switch off and on A parameter is out of range. The parameter could not
be replaced by its default value. The error will dis-
appear at the next power-on Active Active
42 P42 -
Current output
Check cables Current loop is disconnected or the loop resistance
is too big Active Active
43 P43 -
Internal error
Switch off and on Too many errors occured at the same time
Some errors are not detected correctly Active Active
44 P44 -
CT SENSORPROM
â
SENSORPROM
â
unit has been used as CT version Active Active
60 F60 -
CAN comm. error
Converter/AOM CAN bus communication error. An add-on module, the
display module or the converter is defect Zero Inactive
61 F61 -
SENSORPROM
â
error
Replace It is not possible to rely on the data in SENSOR-
PROM
â
unit anymore Active Active
62 F62 -
SENSORPROM
â
ID
Replace The SENSORPROM
â
unit ID does not comply with the
product ID.
The SENSORPROM
â
unit is from another type of
product MASSFLO
â
, SONOFLO
â
etc. Zero Inactive
63 F63 -
SENSORPROM
â
Replace It is not possible to read from the SENSORPROM
â
unit anymore. Active Active
70 F70 -
Coil current
Check cables Coil excitation has failed Active Active
71 F71 -
Internal error
Replace converter Internal convertion error in ASIC Active Active
8.7.2
List of error numbers

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Danfoss MAGFLO MAG 5000 and is the answer not in the manual?

Danfoss MAGFLO MAG 5000 Specifications

General IconGeneral
BrandDanfoss
ModelMAGFLO MAG 5000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals