5-8 Magellan
®
8300/8400
Shift Test
The Shift Test checks to ensure that items placed anywhere on the weigh-
ing surface of the scanner/scale are weighed properly. Refer to
Figure 5-2
when performing this test.
1. Place and remove in succession, a fifteen-pound load on the center
of each of the four quadrants (A, B, C, and D in Figure 5-2) of the
scanner/scales weigh platter. Verify that a reading of between 14.99
and 15.01 lb registers under load and that the display returns to
0.00 between each load.
2. After verifying the accuracy of each quadrant of the weighing sur-
face, remove all weight from the scale. You have completed the Shift
Te st .
Figure 5-2. Shift Test
A
C
B
D
AC
B
D

Table of Contents

Other manuals for Datalogic MAGELLAN 8300

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Datalogic MAGELLAN 8300 and is the answer not in the manual?

Datalogic MAGELLAN 8300 Specifications

General IconGeneral
Read rate (max)3900 reads/s
Decoding standards1D, GS1 DataBar
Standard interfacesKeyboard Wedge, RS-232
USB 2.0 ports quantity1
Input voltage100 - 250 V
Product colorMetallic
Storage temperature (T-T)-40 - 70 °C
Operating temperature (T-T)10 - 40 °C
Operating relative humidity (H-H)5 - 95 %
Weight and Dimensions IconWeight and Dimensions
Depth102 mm
Width292 mm
Height130 mm

Related product manuals