97 
 
Step  Action 
Displayed 
Message 
Comment(s) 
8 
Use the UP, DOWN, LEFT, and 
RIGHT Buttons to set the Gain 
Number to 00 and then press the 
ENTER Key.  
 
Record the reading as a TOF 
Value for Gain Number 00 in the 
table.  
 
Increment the Gain Number and 
press the ENTER Key, note the 
TOF Value. Repeat this process to 
document the TOF Values for the 
thirty remaining Gain Numbers.  
 
Calibration Table 
Gain 
Number 
Label 
Value 
TOF 
Value 
Difference 
Value 
00 252 248  
01 250 245  
02 248 234  
... ... ... 
 
31 131 14   
 
TRAN SENSOR 
GAIN 
 
 
(0 - 
31) 
YYY
 
 
 00
 
Note: The LEFT and RIGHT Buttons 
shift the cursor position.  
9 
In the table, where both the Label 
Value and TOF Value are above 
20, subtract them and record the 
result as a Difference Value (see 
below).  Then find the largest 
Difference Value in your table – its 
corresponding Gain Number will be 
used for sampling the media. 
 
Calibration Table 
Gain 
Number 
Label 
Value 
TOF 
Value 
Difference 
Value 
00 252 248  4 
01 250 245  5 
... ... ...  ... 
06 200 88  112 
07 189 58  131 
08  178  32  146 
09 167 19  N/A 
... ... ...  ... 
30 135 14  N/A 
31 131 14  N/A 
 
TRAN SENSOR 
GAIN 
 
 
(0 - 
31) 
YYY
 
 
 31
 
In this example, Gain Number 08 is 
chosen because, where both the 
Label Value (178) and TOF Value 
(32) are above twenty, it delivers the 
highest Difference Value (146).