EasyManua.ls Logo

Dexcom SEVEN - Page 28

Dexcom SEVEN
144 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
24 | SEVEN
CONTINUOUS GLUCOSE MONITORING SYSTEM
DOUBLE BLOOD DROP PROMPT: Appears 2 hours
after you insert a new Sensor and indicates that you
need to take 2 fingerstick BG readings on your BG meter
and enter them to get into calibration.
SHADEDOUT BLOOD DROP PROMPT: Notifies you
that only 1 fingerstick BG reading has been successfully
entered from the BG meter to the Receiver. You need
to take and upload at least 1 more BG reading for
calibration.
CALIBRATION UPDATE SCREEN: Indicates it is time
for you to take a calibration fingerstick on the BG meter
and enter it to the Receiver.
FAILED SENSOR SCREEN: Indicates the Sensor is not
working properly. You will need to replace your Sensor
at this time.
NO COMMUNICATION: Indicates that the Receiver
and Transmitter are not “talking” and glucose data has
not been sent in the last 5 minutes. See Section 4.6 for
more information.
AWAITING CALIBRATION FINGERSTICK: Indicates
a fingerstick BG reading is needed for calibration or the
Receiver is processing a current fingerstick BG reading.
See Section 3.4 for Calibration Troubleshooting tips.

Table of Contents

Related product manuals