EasyManua.ls Logo

Dexcom SEVEN - Page 56

Dexcom SEVEN
144 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
52 | SEVEN
CONTINUOUS GLUCOSE MONITORING SYSTEM
e 1-Hour Trend Screen with a Blood Drop in the upper right hand corner (“Status Box”) 7.
shown below will appear once the upload is complete. e Blood Drop symbol will be
replaced by a glucose reading once the Sensor successfully calibrates.
Press the Push button on the connection cable to disconnect the cable from the BG 8.
meter and the Receiver (the order that you disconnect the cable does not matter).
A glucose reading will appear on the Receiver in about 15 minutes, and glucose 9.
readings will be updated every 5 minutes.
If it has been more than 20 minutes and you have not received a reading, see Section 10.
3.4 “Calibration Troubleshooting.
NOTE
All fingerstick readings taken with your OneTouch Ultra meter are evaluated for •
calibration. Do not share your blood glucose meter with other people.
It is important that you follow these steps for Auto Upload in the exact order above
•
for successful BG reading transfer and calibration of the Seven System.
HELPFUL HINTS:
Always check that the Receiver and BG meter date and time are the same. •
If you use more than 1 OneTouch Ultra meter during Seven System use, any BG •
fingersticks you take using the non-registered” meter will not be used for calibration
even if you attach this meter with the connection cable. You will have to register
the meter first and then take BG readings with the new meter for Auto-Upload
calibration.