100 Instructions for use Babyleo TN500 SW 1.0n
Operation
5 If oxygen has been used, air the device thoroughly: Open the hood and the side
access panels for at least 3 minutes. Close them again after the airing.
The Start button changes from gray to green when the following conditions
are met:
– The warm-up time has elapsed.
– The Oxygen therapy control is at Off.
– The device has been adequately aired.
If the Start button is displayed in gray, repeat steps 3 to 5.
6 Touch the Start button (3).
7 Press the rotary knob to confirm.
The adjustment takes approximately 1 minute.
A notification will be displayed if the adjustment was successful.
✓ The O
2 sensors have been adjusted.
After the O2 sensors have been adjusted, the date of the next adjustment is
displayed in the Oxygen dialog window. An alarm message will be displayed if
the O
2 sensors have not been adjusted by this date. Additionally, the oxygen
measurement on the main screen is shown in gray. A gray measurement
indicates that the measurement accuracy is no longer guaranteed.
6.13 Heated mattress (option)
This chapter describes how to connect the heated mattress (option) to the device
and how to set the mattress core temperature.
6.13.1 Heated mattress overview
The heated mattress can be placed on the mattress tray of the device as an
alternative to the SoftBed mattress. The heated mattress provides additional
warmth for the patient by means of thermal conduction. The temperature of the
heated mattress can be set manually or automatically.
The following table shows the setting options:
The main screen of the user interface displays the mattress core temperature. The
mattress core temperature is measured inside the mattress. The mattress core
temperature is the mean of 2 temperatures measured at the edges of the heating
element. The mattress core temperature can also be influenced by other heat
sources.
Setting Meaning
35 to 39 °C
(95 to 102 °F)
Setting range for the mattress core temperature
The mattress core temperature can be set in steps of 0.5 °C
(0.5 °F).
Auto – At a lower air temperature, a lower mattress core tem-
perature is selected automatically.
– At a higher air temperature, a higher mattress core tem-
perature is selected automatically.
Off The heated mattress is deactivated.