Instructions for use Babyleo TN500 SW 1.0n 189
Technical data
Side wall height >115 mm (mattress tray in top position),
>205 mm (mattress tray in bottom position)
Weight <140 kg (308.6 lbs) (without options and acces-
sories)
Height adjustment of the device
Height adjustment range 400 mm (15.75 in)
Operating time 2 minute continuous operation with subse-
quent interruption for 18 minutes to cool the
drive
Inclination of the mattress tray
Bed-tilt mechanism Continuous up to 13° inclination on both sides
Electrical protection class
Device Class I
Skin temperature sensor Type BF
Scale (option) Type BF
Heated mattress (option) Type BF
Classification in compliance with Directive
93/42/EEC, Annex IX
Class IIb
UMDNS code
Universal Medical Device Nomenclature
System - nomenclature for medical devices
15-610
Electromagnetic compatibility (EMC) Tested in compliance with IEC 60601-1-2
Standards
Applicable standards IEC 60601-1
IEC 60601-2-19, for incubators
IEC 60601-2-21, for radiant warmers
OIML-R76-1 and
EN 45501, for scales
14.3 Performance characteristics
Operating noise
Operating volume in patient compartment Typically 40 dB(A)
Measured without oxygen application
Alarm volume
Setting range for high- and medium-priority
alarms
50 dB(A) to 70 dB(A)
Measured in accordance with IEC 60601-2-19
and IEC 60601-2-21
Power failure alarm
Alarm volume 65 dB(A)
Duration 10 minutes
14.2 Operating data (continued)