116 Instructions for use Babyleo TN500 SW 1.0n
Operation
6.18 Scale (option)
This chapter describes how to use the scale (option) in the device.
6.18.1 Overview of the scale
The scale is integrated in the bed support. When weighing the patient, the objects
resting on the mattress are also weighed. In order to precisely determine the
patient's weight, the scale must be tared. In order to tare the scale, the patient must
be lifted up so that the weight of the mattress tray and the objects resting on it can
be determined. This weight is deducted from the measured weight when the patient
is subsequently weighed so that only the patient's weight is displayed. An acoustic
signal sounds when the scale is tared and also when the weighing process is
complete. The patient can be weighed during incubator operation and radiant
warmer operation.
Fig. 64 Scale dialog window
28493
No. Designation Description
1 Last weight Shows the last measured weight.
2 Weigh patient Activates the weighing process.
3 Weight (NET) [g] Shows the weight after taring.
4 Adjusted weight The symbol indicates whether the measured
weight was adjusted by the user.
1
2
4 3