Instructions for use Babyleo TN500 SW 1.0n 77
Operation
6.8 Skin temperature mode
This chapter describes skin temperature mode.
6.8.1 Overview of skin temperature mode
Skin temperature mode can be activated in incubator operation and in radiant
warmer operation. In skin temperature mode, the temperature of the device is
regulated by means of the set value for the skin temperature. This mode requires
the use of skin temperature sensors. When the skin temperature sensors are
connected, the user sets the desired skin temperature on the device. The
temperature can be set in steps of 0.1 °C (0.1 °F).
The following table shows the temperature ranges:
To set the temperature in the extended range, first confirm the extended range
by pressing the rotary knob.
In incubator operation the air temperature in the device is regulated according to
the set skin temperature. In radiant warmer operation the radiant warmer power is
regulated according to the set skin temperature. If the current skin temperature
differs from the target value, an alarm is triggered. Additional alarm limits can be set
by means of the Advanced skin temperature monitoring function. If the central
skin temperature and the peripheral skin temperature are measured, changes in
skin temperature can be detected more quickly using the Thermomonitoring
function. For further information, see the following
section: ''Thermomonitoring'' (page 83)
To open the quick setup dialog window, touch a skin temperature value. The
buttons for the alarm limits are displayed.
Setting range Extended range
34 to 37 °C (93.2 to 98.6 °F) 37.1 to 38 °C (98.7 to 100.4 °F)